久草精品视频在线播放,中文人妻AV久久人妻水密桃,av免费观看精品,亚洲五月激情综合图片区 日本成熟少妇A∨网站,久久看亚洲欧美,久久久久久人妻精品一区,久久亚洲Aⅴ无码影院

產(chǎn)品中心

一步法小鼠新喋呤(NPT)酶聯(lián)免疫吸附檢測試劑盒
一鍵復(fù)制產(chǎn)品信息
ELK100ES
規(guī)格: 價格:
48T ¥2240.00
96T ¥3200.00

Overview

Product name: EasyStep Mouse NPT(Neopterin) ELISA Kit
Reactivity: Mouse
Alternative Names: Neopterin
Assay Type: Competitive Inhibition
Sensitivity: 0.4 nmol/L
Standard: 30 nmol/L
Detection Range: 0.469-30 nmol/L
Sample Type: Serum, plasma
Assay Length: 1.5h
Research Area: Cytokine;Infection immunity;
Test principle: This assay employs the competitive inhibition enzyme immunoassay technique. The microtiter plate provided in this kit has been pre-coated with Mouse NPT antigen, and the Mouse NPT standard plate wells that pre-coated using protein-related techniques are provided separately. Standard/Sample Diluent Buffer or samples are added to the appropriate microtiter plate wells ,then added a HRP-conjugated antibody specific to Mouse NPT to each microplate well and incubated . After TMB substrate solution is added, the enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ±10nm. The concentration of Mouse NPT in the samples is then determined by comparing the OD of the samples to the standard curve.

標(biāo)準(zhǔn)曲線

Concentration (nmol/L) OD Corrected OD
30.00 0.215 0.000
15.00 0.352 0.000
7.50 0.567 0.000
3.75 0.833 0.000
1.88 1.157 0.000
0.94 1.585 0.000
0.47 1.715 0.000
0.00 2.226 0.000

精密度

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

回收率

Matrices listed below were spiked with certain level of recombinant NPT and the recovery rates were calculated by comparing the measured value to the expected amount of NPT in samples.
Matrix Recovery range Average
serum(n=5) 84-98% 91%
EDTA plasma(n=5) 80-96% 88%
Heparin plasma(n=5) 82-98% 90%

線性

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of NPT and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 89-99% 83-95% 88-94% 96-101%
EDTA plasma(n=5) 85-97% 90-99% 82-101% 93-105%
Heparin plasma(n=5) 85-97% 87-96% 82-95% 92-105%